Dosimetry - RAD 60 / Alarming Radiation Dosimeter

Description

The RADOS RAD-60 Personal Alarm Dosimeter is a precise and reliable instrument for ensuring the personal safety of the user. Ideally, the RAD-60 is used in everyday radiation monitoring, in stand-alone conditions. If your needs grow for a more sophisticated system, the versatile RAD-60 can also be integrated into an Access Control System. The RAD-60 can be switched into System Mode, for the purpose of tracking Personnel Dose records and generating compliance reports.
The design includes state-of-the-art technology with built in memory for retrieving dose, even during power-down. It eliminates outside interference from shock and RF. The RAD-60 is easily programmed by the user, has a digital display, and operates with a single AAA alkaline battery.

With the push of a button, you can turn the unit on/off, change the digital display to read dose or dose rate, select from several dose and dose rate alarm levels, turn the chirp functionon/off, reset the integrated dose, and perform battery tests. The large digital display gives instant dose or dose rate readings and the audible alarm is loud (>85dba)

Detector

Energy compensated Si-Diode

Button Functions
Front panel button has the following functions:
change display priority (dose/dose rate)
switch ON/OFF
chirp ON/OFF
reset integrated dose
change alarm thresholds
activate battery test

Measurement Range
Dose: 1 uSv - 9.99 Sv or 0.1 mrem - 999 rem

Energy Response
Hp(10), 60 keV - 3 MeV, better than + or - 25%, up to 6 MeV, better than + or - 35%

Dose Rate
5 uSv/h - 3 Sv/h or 0.5 mrem/h - 300 rem/h

Dose Rate Linearity
Better than + or - 15% up to 3 Sv/h (300 rem/h)

Calibration
Better than + or -5% (Cs-137, 662 keV at 2 mSv/h), Hp(10)

Audible Alarms
Seven separate alarms, sound level typically better than 85 dBA at 30 cm
integrated dose
dose rate
dose overflow
dose rate overflow at 3 Sv/h or 300 rem/h
low battery 1 and 2
defect

Alarm Thresholds
Six preset values each for integrated dose and dose rate-push button selection

Connectors
Infrared communication via bottom of the dosimeter

Power Requirements
One triple A alkaline cell, typical life is 1800 hours in background (dose mode)

Temperature Range
-20 - + 50° C operational, humidity up to 90% RH, non-condensed -20 - + 70° C storing

Weight
80 g (including battery)

Size
78 x 67 x 22 mm

Warranty
1 Year Limited