Digital Detectors - Digilert 200
Description
Specifications
Detector
Halogen-quenched GM tube with thin mica end window. Mica window density 1.5-2.0 mg/cm2. Effective window diameter is .360 inch. Side wall is .012 inch thick.
Operating Range
mR/hr - .001 (1µR) to 200 mR/hr
µSv/hr - .01 to 2000
CPM - 0 to 214,000
CPS - 0 to 3575
Total Counts - 1 to 9,999,000 counts
Accuracy
Typically ±15% from factory. , ±10% (NIST).
Energy Sensitivity
1070 CPM/mR/hr referenced to Cs137
Detects alpha down to 2.5 MeV; typical detection efficiency at 3.6 MeV is greater than 80%.
Detects beta at 50 keV with typical 35% detection efficiency.
Detects beta at 150 keV with typical 75% detection efficiency.
Detects gamma and x-rays down to 10 keV typical through the window, 40 keV minimum through the case.
Anti-Saturation
Readout will OVERRANGE in radiation fields as high as 100 times the maximum reading.
Alert
Pulsating beeper sounds the alert. Adjustable alert levels are used for mR/hr / CPM, and µSv/hr / CPS. Alarm will sound when in Timer Mode when set alarm threshold is reached.
Display
Backlit 4-digit liquid crystal display with mode indicators. Display updates every 3 seconds.
Count Light
Red LED flashes with each count.
Audio Indicator
Sounds with each count (can be switched off for silent operation)
Outputs
3.5mm jack for CMOS or TTL devices and counts to computer or datalogger. USB for use with Observer USB Software for PCs
Power Requirements
One 9-volt alkaline battery. Battery life is approx. 1200 hours at normal background radiation levels.
Temperature Range
-10° to +50°C (14° to 122°F)
Weight
217 g (7.7 oz.)
Size
150 x 80 x 31 mm (6.1 x 3.1 x 1.2 in.)
Includes
Carrying Case, Observer USB Software Download, USB Cable, Certification of Conformance.
Options
Xtreme Boot & Stand
Limited Warranty
1 year limited warranty